User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Kyowa DC Cal Signal Sub DATE: 27-Aug-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 25 NUMBER OF LINES: 363 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 34401A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD { DC Calibration Signals } 1.002 DISP Connect the 5700A HI/LO OUTPUT to the UUT 1.002 DISP CH 1 RECORD BNC. 1.003 DISP Connect the UUT CH 1 REPROCUCE OUTPUT BNC to the 1.003 DISP HI/LO INPUT on the 34401A. 1.004 DISP In the UUT FUNCTION group press the MAN button. 1.004 DISP In the UUT ENTRY keypad press .2, EXC. 1.004 DISP Press MAN in the UUT FUNCTION group. 1.005 DISP In the UUT FUNCTION group press CAL SET. 1.005 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 1.005 DISP .2, EXC. 1.006 DISP Press CAL SET in the FUNCTION group. 1.007 5700 0V S 2W 1.008 IEEE [@34401]FUNC "VOLT:DC"[GTL] 1.009 DISP Adjust the UUT CH 1 ZERO control for a reading 1.009 DISP of 0.00mV on the DVM. 1.010 5700 0.2V S 2W 1.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 1.012 DISP Adjust the UUT CH 1 GAIN control for a reading 1.013 DISP of 5.000V on the DVM. 1.014 DISP Press the CAL button 2 times. 1.015 IEEE [@34401]FUNC "VOLT:DC" 1.016 IEEE [@34401]READ?[I] 1.017 ACC .2000V 0.0045% 1.018 MATH MEM = MEM / 5 1.019 MATH MEM = .2 * MEM 1.020 MEMC V 1% 2.001 DISP Press the CAL button 1 time. 2.002 IEEE [@34401]FUNC "VOLT:DC" 2.003 IEEE [@34401]READ?[I] 2.004 ACC -.2000V 0.0045% 2.005 MATH MEM = MEM / -5 2.006 MATH MEM = -.2 * MEM 2.007 MEMC V 1% 3.001 DISP Press CAL 1 time (CAL LED extinguished). 3.002 DISP In the UUT FUNCTION group press the MAN button. 3.002 DISP In the UUT ENTRY keypad press .4, EXC. 3.002 DISP Press MAN in the UUT FUNCTION group. 3.003 DISP In the UUT FUNCTION group press CAL SET. 3.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 3.004 DISP .4, EXC. 3.005 DISP Press CAL SET in the FUNCTION group. 3.006 5700 0V S 2W 3.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 3.008 DISP Adjust the UUT CH 1 ZERO control for a reading 3.009 DISP of 0.00mV on the DVM. 3.010 5700 0.4V S 2W 3.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 3.012 DISP Adjust the UUT CH 1 GAIN control for a reading 3.013 DISP of 5.000V on the DVM. 3.014 DISP Press the CAL button 2 times. 3.015 IEEE [@34401]FUNC "VOLT:DC" 3.016 IEEE [@34401]READ?[I] 3.017 ACC .4000V 0.0045% 3.018 MATH MEM = MEM / 5 3.019 MATH MEM = .4 * MEM 3.020 MEMC V 1% 4.001 DISP Press the CAL button 1 time. 4.002 IEEE [@34401]FUNC "VOLT:DC" 4.003 IEEE [@34401]READ?[I] 4.004 ACC -.4000V 0.0045% 4.005 MATH MEM = MEM / -5 4.006 MATH MEM = -.4 * MEM 4.007 MEMC V 1% 5.001 DISP Press CAL 1 time (CAL LED extinguished). 5.002 DISP In the UUT FUNCTION group press the MAN button. 5.002 DISP In the UUT ENTRY keypad press .6, EXC. 5.002 DISP Press MAN in the UUT FUNCTION group. 5.003 DISP In the UUT FUNCTION group press CAL SET. 5.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 5.004 DISP .6, EXC. 5.005 DISP Press CAL SET in the FUNCTION group. 5.006 5700 0V S 2W 5.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 5.008 DISP Adjust the UUT CH 1 ZERO control for a reading 5.009 DISP of 0.00mV on the DVM. 5.010 5700 0.6V S 2W 5.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 5.012 DISP Adjust the UUT CH 1 GAIN control for a reading 5.013 DISP of 5.000V on the DVM. 5.014 DISP Press the CAL button 2 times. 5.015 IEEE [@34401]FUNC "VOLT:DC" 5.016 IEEE [@34401]READ?[I] 5.017 ACC .6000V 0.0045% 5.018 MATH MEM = MEM / 5 5.019 MATH MEM = .6 * MEM 5.020 MEMC V 1% 6.001 DISP Press the CAL button 1 time. 6.002 IEEE [@34401]FUNC "VOLT:DC" 6.003 IEEE [@34401]READ?[I] 6.004 ACC -.6000V 0.0045% 6.005 MATH MEM = MEM / -5 6.006 MATH MEM = -.6 * MEM 6.007 MEMC V 1% 7.001 DISP Press CAL 1 time (CAL LED extinguished). 7.002 DISP In the UUT FUNCTION group press the MAN button. 7.002 DISP In the UUT ENTRY keypad press .8, EXC. 7.002 DISP Press MAN in the UUT FUNCTION group. 7.003 DISP In the UUT FUNCTION group press CAL SET. 7.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 7.004 DISP .8, EXC. 7.005 DISP Press CAL SET in the FUNCTION group. 7.006 5700 0V S 2W 7.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 7.008 DISP Adjust the UUT CH 1 ZERO control for a reading 7.009 DISP of 0.00mV on the DVM. 7.010 5700 0.8V S 2W 7.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 7.012 DISP Adjust the UUT CH 1 GAIN control for a reading 7.013 DISP of 5.000V on the DVM. 7.014 DISP Press the CAL button 2 times. 7.015 IEEE [@34401]FUNC "VOLT:DC" 7.016 IEEE [@34401]READ?[I] 7.017 ACC .8000V 0.0045% 7.018 MATH MEM = MEM / 5 7.019 MATH MEM = .8 * MEM 7.020 MEMC V 1% 8.001 DISP Press the CAL button 1 time. 8.002 IEEE [@34401]FUNC "VOLT:DC" 8.003 IEEE [@34401]READ?[I] 8.004 ACC -.8000V 0.0045% 8.005 MATH MEM = MEM / -5 8.006 MATH MEM = -.8 * MEM 8.007 MEMC V 1% 9.001 DISP Press CAL 1 time (CAL LED extinguished). 9.002 DISP In the UUT FUNCTION group press the MAN button. 9.002 DISP In the UUT ENTRY keypad press 1, EXC. 9.002 DISP Press MAN in the UUT FUNCTION group. 9.003 DISP In the UUT FUNCTION group press CAL SET. 9.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 9.004 DISP 1, EXC. 9.005 DISP Press CAL SET in the FUNCTION group. 9.006 5700 0V S 2W 9.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 9.008 DISP Adjust the UUT CH 1 ZERO control for a reading 9.009 DISP of 0.00mV on the DVM. 9.010 5700 1V S 2W 9.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 9.012 DISP Adjust the UUT CH 1 GAIN control for a reading 9.013 DISP of 5.000V on the DVM. 9.014 DISP Press the CAL button 2 times. 9.015 IEEE [@34401]FUNC "VOLT:DC" 9.016 IEEE [@34401]READ?[I] 9.017 ACC 1.000V 0.0045% 9.018 MATH MEM = MEM / 5 9.019 MATH MEM = 1 * MEM 9.020 MEMC V 1% 10.001 DISP Press the CAL button 1 time. 10.002 IEEE [@34401]FUNC "VOLT:DC" 10.003 IEEE [@34401]READ?[I] 10.004 ACC -1.000V 0.0045% 10.005 MATH MEM = MEM / -5 10.006 MATH MEM = -1 * MEM 10.007 MEMC V 1% 11.001 DISP Press CAL 1 time (CAL LED extinguished). 11.002 DISP In the UUT FUNCTION group press the MAN button. 11.002 DISP In the UUT ENTRY keypad press 1.2, EXC. 11.002 DISP Press MAN in the UUT FUNCTION group. 11.003 DISP In the UUT FUNCTION group press CAL SET. 11.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 11.004 DISP 1.2, EXC. 11.005 DISP Press CAL SET in the FUNCTION group. 11.006 5700 0V S 2W 11.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 11.008 DISP Adjust the UUT CH 1 ZERO control for a reading 11.009 DISP of 0.00mV on the DVM. 11.010 5700 1.2V S 2W 11.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 11.012 DISP Adjust the UUT CH 1 GAIN control for a reading 11.013 DISP of 5.000V on the DVM. 11.014 DISP Press the CAL button 2 times. 11.015 IEEE [@34401]FUNC "VOLT:DC" 11.016 IEEE [@34401]READ?[I] 11.017 ACC 1.200V 0.0045% 11.018 MATH MEM = MEM / 5 11.019 MATH MEM = 1.2 * MEM 11.020 MEMC V 1% 12.001 DISP Press the CAL button 1 time. 12.002 IEEE [@34401]FUNC "VOLT:DC" 12.003 IEEE [@34401]READ?[I] 12.004 ACC -1.200V 0.0045% 12.005 MATH MEM = MEM / -5 12.006 MATH MEM = -1.2 * MEM 12.007 MEMC V 1% 13.001 DISP Press CAL 1 time (CAL LED extinguished). 13.002 DISP In the UUT FUNCTION group press the MAN button. 13.002 DISP In the UUT ENTRY keypad press 1.4, EXC. 13.002 DISP Press MAN in the UUT FUNCTION group. 13.003 DISP In the UUT FUNCTION group press CAL SET. 13.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 13.004 DISP 1.4, EXC. 13.005 DISP Press CAL SET in the FUNCTION group. 13.006 5700 0V S 2W 13.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 13.008 DISP Adjust the UUT CH 1 ZERO control for a reading 13.009 DISP of 0.00mV on the DVM. 13.010 5700 1.4V S 2W 13.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 13.012 DISP Adjust the UUT CH 1 GAIN control for a reading 13.013 DISP of 5.000V on the DVM. 13.014 DISP Press the CAL button 2 times. 13.015 IEEE [@34401]FUNC "VOLT:DC" 13.016 IEEE [@34401]READ?[I] 13.017 ACC 1.400V 0.0045% 13.018 MATH MEM = MEM / 5 13.019 MATH MEM = 1.4 * MEM 13.020 MEMC V 1% 14.001 DISP Press the CAL button 1 time. 14.002 IEEE [@34401]FUNC "VOLT:DC" 14.003 IEEE [@34401]READ?[I] 14.004 ACC -1.400V 0.0045% 14.005 MATH MEM = MEM / -5 14.006 MATH MEM = -1.4 * MEM 14.007 MEMC V 1% 15.001 DISP Press CAL 1 time (CAL LED extinguished). 15.002 DISP In the UUT FUNCTION group press the MAN button. 15.002 DISP In the UUT ENTRY keypad press 1.6, EXC. 15.002 DISP Press MAN in the UUT FUNCTION group. 15.003 DISP In the UUT FUNCTION group press CAL SET. 15.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 15.004 DISP 1.6, EXC. 15.005 DISP Press CAL SET in the FUNCTION group. 15.006 DISP Press the CAL button 2 times. 15.007 IEEE [@34401]FUNC "VOLT:DC" 15.008 IEEE [@34401]READ?[I] 15.009 ACC 1.600V 0.0045% 15.010 MATH MEM = MEM / 5 15.011 MATH MEM = 1.6 * MEM 15.012 MEMC V 1% 16.001 DISP Press the CAL button 1 time. 16.002 IEEE [@34401]FUNC "VOLT:DC" 16.003 IEEE [@34401]READ?[I] 16.004 ACC -1.600V 0.0045% 16.005 MATH MEM = MEM / -5 16.006 MATH MEM = -1.6 * MEM 16.007 MEMC V 1% 17.001 DISP Press CAL 1 time (CAL LED extinguished). 17.002 DISP In the UUT FUNCTION group press the MAN button. 17.002 DISP In the UUT ENTRY keypad press 2, EXC. 17.002 DISP Press MAN in the UUT FUNCTION group. 17.003 DISP In the UUT FUNCTION group press CAL SET. 17.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 17.004 DISP 2, EXC. 17.005 DISP Press CAL SET in the FUNCTION group. 17.006 5700 0V S 2W 17.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 17.008 DISP Adjust the UUT CH 1 ZERO control for a reading 17.009 DISP of 0.00mV on the DVM. 17.010 5700 2V S 2W 17.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 17.012 DISP Adjust the UUT CH 1 GAIN control for a reading 17.013 DISP of 5.000V on the DVM. 17.014 DISP Press the CAL button 2 times. 17.015 IEEE [@34401]FUNC "VOLT:DC" 17.016 IEEE [@34401]READ?[I] 17.017 ACC 2.000V 0.0045% 17.018 MATH MEM = MEM / 5 17.019 MATH MEM = 2 * MEM 17.020 MEMC V 1% 18.001 DISP Press the CAL button 1 time. 18.002 IEEE [@34401]FUNC "VOLT:DC" 18.003 IEEE [@34401]READ?[I] 18.004 ACC 2.000V 0.0045% 18.005 MATH MEM = MEM / -5 18.006 MATH MEM = 2 * MEM 18.007 MEMC V 1% 19.001 DISP Press CAL 1 time (CAL LED extinguished). 19.002 DISP In the UUT FUNCTION group press the MAN button. 19.002 DISP In the UUT ENTRY keypad press 4, EXC. 19.002 DISP Press MAN in the UUT FUNCTION group. 19.003 DISP In the UUT FUNCTION group press CAL SET. 19.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 19.004 DISP 4, EXC. 19.005 DISP Press CAL SET in the FUNCTION group. 19.006 DISP Press the CAL button 2 times. 19.007 IEEE [@34401]FUNC "VOLT:DC" 19.008 IEEE [@34401]READ?[I] 19.009 ACC 4.000V 0.0045% 19.010 MATH MEM = MEM / 5 19.011 MATH MEM = 4 * MEM 19.012 MEMC V 1% 20.001 DISP Press the CAL button 1 time. 20.002 IEEE [@34401]FUNC "VOLT:DC" 20.003 IEEE [@34401]READ?[I] 20.004 ACC -4.000V 0.0045% 20.005 MATH MEM = MEM / -5 20.006 MATH MEM = -4 * MEM 20.007 MEMC V 1% 21.001 DISP Press CAL 1 time (CAL LED extinguished). 21.002 DISP In the UUT FUNCTION group press the MAN button. 21.002 DISP In the UUT ENTRY keypad press 6, EXC. 21.002 DISP Press MAN in the UUT FUNCTION group. 21.003 DISP In the UUT FUNCTION group press CAL SET. 21.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 21.004 DISP 6, EXC. 21.005 DISP Press CAL SET in the FUNCTION group. 21.006 5700 0V S 2W 21.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 21.008 DISP Adjust the UUT CH 1 ZERO control for a reading 21.009 DISP of 0.00mV on the DVM. 21.010 5700 6V S 2W 21.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 21.012 DISP Adjust the UUT CH 1 GAIN control for a reading 21.013 DISP of 5.000V on the DVM. 21.014 DISP Press the CAL button 2 times. 21.015 IEEE [@34401]FUNC "VOLT:DC" 21.016 IEEE [@34401]READ?[I] 21.017 ACC 6.000V 0.0045% 21.018 MATH MEM = MEM / 5 21.019 MATH MEM = 6 * MEM 21.020 MEMC V 1% 22.001 DISP Press the CAL button 1 time. 22.002 IEEE [@34401]FUNC "VOLT:DC" 22.003 IEEE [@34401]READ?[I] 22.004 ACC -6.000V 0.0045% 22.005 MATH MEM = MEM / -5 22.006 MATH MEM = -6 * MEM 22.007 MEMC V 1% 23.001 DISP Press CAL 1 time (CAL LED extinguished). 23.002 DISP In the UUT FUNCTION group press the MAN button. 23.002 DISP In the UUT ENTRY keypad press 8, EXC. 23.002 DISP Press MAN in the UUT FUNCTION group. 23.003 DISP In the UUT FUNCTION group press CAL SET. 23.004 DISP In the UUT ENTRY keypad press SFT1, CALA, EXC, SFT1, DC, 23.004 DISP 8, EXC. 23.005 DISP Press CAL SET in the FUNCTION group. 23.006 5700 0V S 2W 23.007 IEEE [@34401]FUNC "VOLT:DC"[GTL] 23.008 DISP Adjust the UUT CH 1 ZERO control for a reading 23.009 DISP of 0.00mV on the DVM. 23.010 5700 8V S 2W 23.011 IEEE [@34401]FUNC "VOLT:DC"[GTL] 23.012 DISP Adjust the UUT CH 1 GAIN control for a reading 23.013 DISP of 5.000V on the DVM. 23.014 DISP Press the CAL button 2 times. 23.015 IEEE [@34401]FUNC "VOLT:DC" 23.016 IEEE [@34401]READ?[I] 23.017 ACC 8.000V 0.0045% 23.018 MATH MEM = MEM / 5 23.019 MATH MEM = 8 * MEM 23.020 MEMC V 1% 24.001 DISP Press the CAL button 1 time. 24.002 IEEE [@34401]FUNC "VOLT:DC" 24.003 IEEE [@34401]READ?[I] 24.004 ACC -8.000V 0.0045% 24.005 MATH MEM = MEM / -5 24.006 MATH MEM = -8 * MEM 24.007 MEMC V 1% 25.001 DISP Press CAL 1 time (CAL LED extinguished)y. 25.002 END